Title of Invention

PHOTOELECTROCHEMICAL DETERMINATION OF CHEMICAL OXYGEN DEMAND

Abstract A method for determining chemical oxygen demand of a water sample comprises the steps of (a) applying a constant potential bias to a photoelectrochemical cell, having a photoactive working electrode (e.g. a layer of titanium dioxide nanoparticles coated on an inert conductive substrate) and a counter electrode, and containing a supporting electrolyte solution; (b) illuminating the working electrode with a light source and recording the background photocurrent produced at the working electrode from the supporting electrolyte solution; (c) adding a water sample, to be analysed, to the photoelectrochemical cell; (d) illuminating the working electrode with a light source and recording the total photoelectrocurrent produced with the sample; (e) determining the chemical oxygen demand according to the type (exhaustive or non-exhaustive) of degradation conditions employed. An apparatus for carrying out the method is also claimed.
Full Text PHOTOELECTROCHEMICAL DETERMINATION OF CHEMICAL OXYGEN DEMAND
Field Of The Invention
This invention relates to a new method for determining oxygen demand of
water using photoelectrochemical cells. In particular, the invention relates to a
i new photoelectrochemical method of determining chemical oxygen demand of
water samples using a titanium dioxide nanoparticulate semiconductive electrode.
Background To The Invention
Nearly all domestic and industrial wastewater effluents contain organic
compounds, which can cause detrimental oxygen depletion (or demand) in
waterways into which the effluents are released. This demand is due largely to the
oxidative biodegradation of organic compounds by naturally occurring
microorganisms, which utilize the organic material as a food source. In this
process, carbon is oxidised to carbon dioxide, while oxygen is consumed and
reduced to water.
Standard analytical methodologies for the determination of aggregate
properties such as oxygen demand in water are biochemical oxygen demand
(BOD) and chemical oxygen demand (COD). BOD involves the use of
heterotrophic microorganisms to oxidise organic material and thus estimate
oxygen demand. COD uses strong oxidising agents, such as dichromate or
permanganate, to oxidise organic material. BOD analysis is carried out over five
days and oxygen demand determined by titration or with an oxygen probe. COD
measures dichromate or permanganate depletion by titration or
spectrophotometry.
Despite their widespread use for estimating oxygen demand, both BOD and
COD methodologies have serious technological limitations. Both methods are time
consuming and very expensive, costing water industries and local authorities in
excess of $1 billion annually worldwide. Other problems with the BOD assay
include: limited linear working range; complicated, time consuming procedures;
and questionable accuracy and reproducibility (the standard method accepts a
relative standard deviation of ±15% for replicate BOD5 analyses). More
importantly, interpretation of BOD results is difficult since the results tend to be
specific to the body of water in question', depend on the pollutants in the sample
solution and the nature of the microbial seed used. In addition, the BOD
methodologies cannot be used to assess the oxygen demand for many heavily
polluted water bodies because of inhibitory and toxic effects of pollutants on the
heterotropic bacteria.
The COD method is more rapid and less variable than the BOD method and
thus preferred for assessing the oxygen demand of organic pollutants in heavily
polluted water bodies. Despite this, the method has several drawbacks in that it is
time consuming, requiring 2-4 hours to reflux samples, and utilises expensive (e.g.
Ag2SO4), corrosive (e.g. concentrated H2SO4) and highly toxic (Hg(ll) and Cr(VI))
reagents. The use of toxic reagents being of particular environmental concern,
leading to the Cr(VI) method being abandoned in Japan.
Titanium(IV) oxide (TiO2) has been extensively used in photooxidation of
organic compounds. T1O2 is non-photocorrosive, non-toxic, inexpensive, relatively
easily synthesised in its highly catalytic nanoparticulate form, and is highly efficient
in photooxidative degradation of organic compounds.
Fox MA and Tien, T, Anal. Chem, (60 1988) 2278-2282 investigated the
development of a photoelectrochemical detector by employing an anodically
formed TiO2 electrode for use in high-pressure liquid chromatography. This
photoelectrochemical detector is reported as being sensitive to oxidisable
organics, such as alcohols. The electrode system developed by Fox et al had low
photocatalytic efficiency of the system and is difficult to use as it cannot
discriminate between the respective currents generated from the oxidation of water
and organic matter.
Brown, G.N., etal., Anal. Chem, 64 (1992) 427-434 investigated the use of
a photoelectrochemical detector by employing a thermally formed TiO2 electrode
for use as a detector in flow injection analysis and liquid chromatography. The
detector was found to be non-selective in its response to a variety of organic
analytes. Brown et al found that the response of the detector varied with
temperature, duration of heating, oxidative atmosphere, etching of the titanium
wire electrode, amount of doping on the TiO2 detector and solvents. Similar to Fox
et al this electrode system had low photocatalytic efficiency and cannot
discriminate between the currents generated from the oxidation of water and
organic matter.
Matthews R.W. et ai., Analytica Chimica Acta 233 (1990) 171-179 (also the
subject of Australian patent 597165) utilised a TiO2 photocatalytic oxidation system
to determine total carbon in water samples, by placing T1O2 into a slurry or
suspension, photooxidising the organic material with in the sample to evolve
carbon dioxide (CO2). The evolved CO2 was measured to predict TOC value of
the sample. Matthews found that the total organic carbon can be estimated from
the total amount of carbon dioxide purged from photocatalytic cell.
Jiang D. et al., J. Photochem & Photobio A: Chemistry 144 (2001) 197-204 also
investigated the photoelectrochemica! behaviour of nanoporous TiO2 film
electrodes in the photooxidation of methanol. Jiang found that the photocurrent
response of the electrode was greatly influenced by applied potential, light
intensity, methanol concentration and pH. A linear relationship was found to exist
between the photocurrent produced through the photo-oxidation of methanol and
the concentration of methanol in the sample. However, as concentration of
methano! increased the migration of photoelectrons across the TiO2 film and
therefore photogenerated charge separation becomes a rate-limiting step, thus
limiting the working range in which the linear relationship between photocurrent
and concentration occurs.
Lee Kyong-Hoon et al., Electroanalysis 12, No 16 (2000) 1334-1338,
investigated the determination of COD using a microfabricated Clark-type oxygen
electrode and TiO2 fine particles suspended in a sample solution under
photocatalytic oxidative degradation conditions. The current generated from the
oxygen electrode under -800mV applied potential was used to indicate the oxygen
concentration change before and after the photooxidation. The change in oxygen
concentration was then correlated to COD value of the sample.
Kim, Yoon-Chang, et a!., Anal. Chem, 72 (2000) 3379-3382; Analytica
Chimica Acta 432 (2001) 59-66 and Anal. Chem, 74 (2002) 3858-3864 all relate to
the determination of COD using a photocatalytic oxidative degradation of organic
compounds at a titanium dioxide particles, in Anal Chem, 2000, Kim et al
investigated the use of translucent poly(tetrafluroethylene) (PTFE) membrane
having fine particles of TIO2 absorbed or entrapped onto the surface of the
membrane in combining with a oxygen electrode as a possible COD sensor. The
immobilised TiO2 particles serve as an oxidation reagent and the analytical
signal was based on the oxygen concentration measurements between the
working and reference oxygen electrodes.
Calibration curves where established using sodium sulfite (Na2SO3), prior to
determining COD of analytes. In this study Kim et al reports that the membrane
sensor did not show good reproducibility.
In Anaiytica Chimica Acta 432 (2001) 59-66, Kim et al invest/gated the use
of titanium dioxide (TiO2) beads in a photochemical column and an oxygen
electrode as the sensor in determining dissolved oxygen from the photocatalytic
oxidation of organic compound and thus the COD value of the analyte.
In Anal. Chem, 74 (2002) 3858-3864Kim et a! investigated the use of
0.6mm TiO2 beads in a quartz tube in the determination of oxygen consumption
from photochemical oxidation of organic compounds and subsequent calculation of
COD values from the difference in the currents recorded at the reference and
working oxygen electrodes.
The methods described by Lee et a! and Kim et ai above all utilise TiO2 as
an oxidative reagent to replace the traditional reagent used in COD such as
chromate salts, with the analytical signal being obtained via two traditional oxygen
electrodes. There are many disadvantages of their method, which makes the
practical application of the method very difficult.
To date the COD assay methodologies of the prior art are indirect in their
analysis methods requiring calibration and often suffer from having low sensitivity,
poor accuracy, narrow linear working ranges and/or difficult to operate. More
importantly, these prior art COD assay methodologies are matrix dependent due to
the low oxidation efficiency, it is an object of this invention to overcome these
shortcomings.
Summary Of The Invention
To this end this invention provides a method of determining chemical
oxygen demand of a water sample, comprising the steps of
a) applying a constant potential bias to a photoelectrochemical ce!J, having
a photoactive working electrode and a counter electrode, and containing a
supporting electrolyte solution;
b) illuminating the working electrode with a light source, preferably a UV
light source, and recording the background photocurrent produced at the working
electrode from the supporting electrolyte solution;
c) adding a water sample, to be analysed, to the photoelectrochemical cell;
d) illuminating the working electrode with a UV light source and recording
the total photocurrent produced;
e) determining the chemical oxygen demand of the water sample
according to the type of degradation conditions employed.
The working electrode may be formed from any nanoparticulate
semiconductive material capable of photooxidation of organic compounds. The
nanoparticulate semiconductive electrode may be selected from titanium dioxide,
niobium pentoxide, strontium titanate, indium trioxide, calcium titanate, tungsten
trioxide, barium titanate, ferric oxide, zinc oxide, potassium tantalate, tin dioxide,
cadmium oxide, hafnium oxide, zirconium oxide, tantalum pentoxide, chromium
trioxide or yttrium trioxide. Preferably the semiconductive electrode is titanium
dioxide. More preferably the semiconductive electrode is formed by laying
nanoparticles of titanium dioxide on an inert substrate, such as conducting glass.
This invention is partly predicated on the insight that the methods described
in the prior art, utilising photoelectrochemical properties of TiO2 nanoparticle
semiconductive electrodes provide for the direct determination of COD. In the
prior art method, the method relies for accuracy on the two oxygen electrodes,
which have to be identical in responding to the oxygen change. In addition, the
prior art method cannot be used for low COD samples due to the insufficient
sensitivity of the method. Low COD content is important in testing water for
suitability in drinking and cleaning applications. By using this technique the
measurement of milliamperes of current allows much greater sensitivity in the low
COD range.
The method of determining the chemical oxygen demand value of a water
sample may be determined under exhaustive degradation conditions, in which all
organics present in the water sample are oxidised. Under exhaustive degradation
conditions the chemical oxygen demand value can be obtained according to the
steps of;
a) integrating both the background photocurrent and total photocurrent
(ji(1)) to give the background charge and total charge, and subtracting the
background charge from the total charge to determine the net charge Qnet;
for the water sample; and
b) calculating the chemical oxygen demand value utilising formula (I);

The method of determining the chemical oxygen demand value of a water
sample may be determined under non-exhaustive degradation conditions, in which
the organics present in the water sample are partially oxidised. Under non-
exhaustive degradation conditions, all photocurrent are measured under the
diffusion controlled conditions and the method of determining oxygen demand
value may further include the steps of;
adding a standard solution, having known organic concentration or chemical
oxygen demand value, to the photoelectrochemical cell containing the water
sample to be analysed;
illuminating working electrode with a UV light source and recording the
limiting photocurrent produced from the partial oxidative degradation of the
standard solution (il(2)); and
determining the chemical oxygen demand value by;
a) calculating the current for the water sample by subtracting the
limiting background photocurrent (iibackgmund)from the limiting photocurrent of the
sample (il(1)) at a predetermined illumination time according to the formula

b) calculating the limiting photocurrent for the standard solution by
subtracting the background photocurrent (ibackground) from the limiting photocurrent
of the sample and standard (il(2)) at the predetermined illumination time according
to the formula:
The method steps for both the exhaustive and non-exhaustive degradation
condition may be repeated as many times as required to analyse all necessary
> water samples.
The potential bias applied to the electrodes is preferably between -0.1V and
+0.5V. More preferably the potential difference is between approximately +0.15V
and +0.35V.
A supporting electrolyte is used to determine the background photocurrent
and to dilute the water sample to be tested. The determination of the background
photocurrent measures the oxidation of water and this can be deducted from the
sample reading to give the photocurrent due to the oxidation of organic material in
the sample. This measurement may be made as a separate measurement to the
sample reading or when conducting an exhaustive degradation the final steady
current after the oxidation is completed is a measure of the background photo
current. The supporting electrolyte may be selected from sodium nitrate, sodium
perchlorate or any other electrolytes that are electrochemically and
photoelectrochemically stable under the experimental conditions and do not
absorb UV radiation in the range being used. The dilution of the samples
enables the method to have a wide linear range while still keeping the test duration
to a relatively small period.
The water sample is preferably illuminated by a light source having a photo
intensity of between between 1 and 100 mWcm-2. More preferably the frequency
of the light source is between approximately 6 - 9,5mWcm-2.
In another aspect the present invention provides a photoelectrochemica)
assay apparatus for determining oxygen demand of a water sample which consists
of
a) a measuring cell for holding a sample to be analysed
b) a photoactive working electrode and a counter electrode disposed in
said cell,
c) a light source, preferably a UV light source, adapted to illuminate the
photoactive working electrode
d) control means to control the illumination of the working electrode, the
applied potential and signal measurement
e) current measuring means to measure the photocurrent at the working
and counter electrodes
f) analysis means to derive a measure of oxygen demand from the
measurements made by the photocurrent measuring means.
Preferably a reference electrode is also located in the measuring cell and the
working electrode is a nanoparticulate semiconductive electrode preferably
titanium dioxide.
In other embodiments the measuring cell may be a stationary cell with different
ceil geometry and volume, or a flow through cell with different cell geometry and
volume, and with a flow rate adjusted to optimise the sensitivity of the
measurements.

Brief Description of ACCOMPANYING Drawings
To assist in understanding the invention preferred embodiments will now be
described with reference to the following figures in which:
Fig. 1. is a schematic illustration of the analytical signal generation for
use in the chemical oxygen demand method of this invention;
Fig. 2 Schematic of the instrumentation of photoelectrochemical detection
system;
Fig. 3. Schematic diagram of the photoelectrochemical batch (stationary)
cell of this invention;
Fig. 4. is a schematic cross section of a thin-layer photoelectrochemical flow
cell according to this invention;
Fig. 5 Schematic diagram of the thin-layer photoelectrochemical flow cell.
Fig. 6. FIA Manifolds for sample and supporting electrolyte injection in an
automatic COD photoelectrochemical detection system.
Fig. 7. Graphical representation of non-exhaustive degradation,
photocurrent/time profiles of supporting electrolyte, sample and
standard solutions.
Fig. 8. Photocurrent response of a solution containing 40µM of potassium
hydrogen phthalate and 0.1M NaNO3. Photo intensity; 9.1mWcm-2;
Applied potential bias: +0.20V vs Ag/AgCI.
Fig. 9 il -C curves for a range of organic compounds, namely;
1 = p-chlorophenol; 2 = potassium hydrogen phthalate;
'3 = methanol; 4 = d-glucose;
5 = malonic acid; 6 = succinic acid;
7 = gfutaric acid; and 8 = glycine.
Photo intensity: 9.1mWcm-2; Applied potential bias: +0.20V vs
Ag/AgCI.
Fig. 10 Photocurrent response of (a) 0.10M NaNO3 and (b) a solution
containing potassium hydrogen phthalate and 0.1M NaNO3. Photo
intensity: 9.1mWcm-2; Applied potential bias: +0.20V vs Ag/AgCI.
Fig. 11. Q -C curves for a range of organic compounds, namely;
1 = p-chlorophenol; 2 = potassium hydrogen phthalate;
3 = methanol; 4 = d-glucose;
5 = malonic acid; 6 = succinic acid;
7 = glutaric acid; and 8 = glycine.
Photo intensity: 9,1mWcm-2; Applied potential bias: +0.20V vs
Ag/AgCl.
Fig 12. Correlation between experimental COD value and standard COD
value, (a) COD standard test solution (KHP); (b) a synthetic COD
sample containing equal molar concentration of all compounds used in
Figure 8.
Fig. 13. Comparison of PECOD and conventional COD method (dichromate)
in the detection of real samples.
Fig. 14 Photoelectrochemical detection of synthetic examples showing peak
height to concentration
Fig.15 Photoelectrochemica! determination of COD vaiue for the synthetic
samples: (a) the quantitative relationship between the peak height
and concentration (µM) of organic compounds, (b) the quantitative
relationship between the peak height and theoretical COD. (c) the
correlation between the PECOD and theoretical COD for the
synthetic COD test samples.
Fig. 16 Pearson correlation between PECOD and conventional COD method
(dichromate).
Detailed Description Of The Invention
The preferred assay method of the invention takes advantage of the highly
efficient photochemical properties of TiO2 nanoparticulate fiim electrodes to
develop a new, rapid, cost-effective assay for the determination of aggregate
organic properties, such as oxygen demand and in particular COD.
This embodiment is directed to a method of determining chemical oxygen
demand of water samples utilising a nanoparticulate TiO2 film electrode. It will be
appreciated by the person skilled in the art that other nanoparticulate
semiconductive film electrode may be utilised in the method without departing from
the essence of the invention.
The assay method of the invention allows for easy quantification of electron
transfer at a TiO2 nanoparticle film electrode during photocatalytic oxidative
degradation of organic material. This approach overcomes many of the current
problems with existing oxygen demand techniques.
The photocatalytic oxidation approach for COD determination utilizes TiO2
particles as photocatalyst to replace the traditional oxidizing agent, e.g. dichromate
and permanganate, illumination of TiO2, with photons whose energy is equal to or
greater than the band-gap energy, will result in promotion of an electron from the
valence band to the conduction band (see Fig. 1). This promotes an electron (e-)
to the conduction band and leaves a positive photohole (h+) in the valence band.
The photohole is one of the most powerful oxidizers due to its high bandgap
potential (+3.2V for anatase). The photocatalysis can lead to stoichiometric
mineralization of organic compounds due to the strong oxidation power of
photoholes.
MATHEMATICAL DERIVATION
The method of determining chemical oxygen demand of water samples,
according to the invention, utilises photoelectrochemical current (or charge)
generated from photoeiectrochemical oxidative degradation of organic compounds
as an analytical signal.
The photocatalytic degradation efficiency at TiO2 depends on the degree of
recombination of photoeiectrons and holes. With traditional TiO2 photocatalysis
systems, this relies on how fast the photoeiectrons and holes are consumed by the
adsorbed species.
A TiO2 nanoparticulate film electrode is used as the working electrode in a
three-electrode photoelectrochemical cell. By applying an appropriate potential
bias to the working electrode, it becomes more favourable for the photoelectron to
be transferred to the working electrode rather than to the adsorbed O2. The
photoeiectrons are subsequently forced to pass into the external circuit and to the
counter electrode, where the reduction of oxygen (or other species) takes place.
The photocurrent (or charge) is monitored and gives a direct measure of the
oxidation of organic compounds. In effect the assay shunts photoeiectrons
through the external circuit to quantify the extent of oxidative degradation.
Separation of the oxidative and reductive half-reactions (Eqn.s 1 and 2,
below) by imposing the electrochemical potential suppresses the recombination of
photoelectrons and holes. As a result, the degradation efficiency is enhanced.
In addition it has been found that the rate of degradation of organic materials is
independent of O2 concentration as the rate of reduction at the counter electrode
will never be the rate-limiting step of the overall degradation process. Thus
overcoming the prior art problem of oxygen reduction being a rate-limiting step in
the photooxidation of organic material.
Quantification of analytical signal
The photoelectrochemical system described above can be used for two different
degradation models - exhaustive and non-exhaustive degradation. With
exhaustive degradation, 100% of the organic analyte in the sample is consumed;
with non-exhaustive degradation, only a small fraction of available analyte is
consumed and its concentration in the bulk solution remains essentially
unchanged. The former is analogous to bulk electrolysis in which all of the anaiyte
is electrolysed and Faraday's Law is used to quantify the concentration by
measuring the charge passed; the latter is analogous to amperometric methods
where the analytical signal (i.e. current) is dependent on the rate of mass transfer
to the electrode surface. In our case, however, the charge/current produced is the
result of photoelectrochemical processes.
When the exhaustive degradation mode! is employed, the charge (Q) can
be measured by the integration of photocurrent within the degradation period. The
analytical principle can be established using Faraday's Law:
Q = idt =nFN = nFVC (3)
where: N = number of moles of analyte in the sample,
n = number of electrons transferred during the photo- electrochemical
degradation,
F = Faraday constant,
V= sample volume; and
C = analyte concentration.
Since TiO2 oxidises organic compounds to the fully oxidised form of carbon
(i.e. CO2), the value n for a given compound will be a constant. Eqn 3 can
therefore be used to quantify the anaiyte concentration.
In principal, analytically useful photocurrents (or charge) can be obtained
from any photo-electrochemically degradable species. The TiO2 system proposed
is capable of oxidising nearly any organic or low redox state inorganic species (e.g.
Fe2+, Cl-, NH4+, NO2-) In this respect, the proposed system can be employed as a
"universal" detector capable of detecting any compounds that can be
photoelectrochemically oxidised at a TiO2 electrode. In combination with an
appropriate separation system (e.g. HPLC), the concentration of individual
analytes can be determined.
In terms of general water quality issues and pollution control, the effect and
quantification of aggregate mixtures of organics (such as in COD or BOD analysis)
are often more important than the analysis of single species. The proposed
photoelectrochemical system is capable of determining such aggregate properties
by summing the charge measured from individual photo-electrochemically
degradable compounds within a mixture (Eqn 4).

The measured charge, Q, is simply the total amount of electron transfer that
results from the degradation of all compounds in the sample. Given that oxidation
by O2 can be represented as:

where one oxygen molecule is equivalent to 4 electrons, the measured Q value
can be easily converted into an equivalent O2 concentration (or oxygen demand)
value:

For exhaustive degradation, the equivalent COD value of the sample can
therefore be represented as:

in the case of non-exhaustive degradation, the quantitative relationship
between the photocurrent and the concentration of the analyte can be developed
using a well-known semi-empirical treatment of Steady-State Mass Transfer [A. J.
Bard and L R. Faulkner, Electrochemical Methods-Fundamental and Applications.
John Wiley & Sons, Inc. New York. 2001], Under conditions of forced convection,
the rate of mass transfer (dN/df) to an electrode is directly proportional to the
concentration gradient at the electrode surface.

where, Cb, - concentration of analyte in the bulk solution;
Cs = concentration of analyte at the electrode surface;
k = mass-transfer coefficient [A. J. Bard and L. R. Faulkner,
Electrochemical Methods-Fundamental and Applications. John Wiley & Sons,
Inc. New York. 2001] which is = D/5, where D = diffusion coefficient and 5 =
thickness of stagnant layer.
When sufficient photo intensity and adequate potential bias are employed,
and the overall process is controlled by mass transfer, then, Cs(x=0) [Cb-Cs(x=0)3˜Cb, The maximum rate of mass transfer, (dN/dt)i is achieved and the
rate of overall reaction equals:

If we again assume that after photochemical oxidation the analyte is fully
oxidised, then the number of electrons transferred (n) during photoelectrochemical
degradation is a constant, for a given analyte. The limiting photocurrent (il) can,
therefore, be used to represent the rate of reaction:

where A = active electrode area.
The development of equation 10, by the current inventors, leads to definition
of the quantitative relationship between the limiting photocurrent and the
concentration of analyte and can be regarded as the principle of analysis.
This analytical principle can again be applied to determine concentrations of
individual analytes (and serve as a "universal" detection system) or to aggregate
mixtures of organics (to determine properties such as COD). Whilst formula 10
above may allow for the general determination of analyte concentration, it is the
application of the findings from formula (10) to the determination of COD that
assists in addressing one or more of the disadvantages in the prior art
methodologies. Standard analytical and mathematical techniques may be used to
calculate the COD of a sample from the limiting photocurrent measured in a
photoelectrochemicai cell utilising a nanoparticulate TiO2 semiconductive electrode
in the manner described in more detail below.
METHOD OF DETERMINING COD
FORMATION OF TiO2 ELECTRODE
A) Synthesis of TiO2 Colloid
A1) A mixture of 12.5ml titanium butoxide and 4ml isopropyl alcohol was added,
drop-wise, to 150ml 0,1 M nitric acid solution under vigorous stirring at room
temperature. After the hydrolysis the slurry was heated to 80°C and stirred for 8h
to achieve peptization. The colloid is then filtered to remove the nonpeptized
agglomerates. For better crystallinity of the nanoparticles, the colloid was
hydrothermally treated in an autoclave at 200°C for 12h. During the autoclaving
sedimentation occurred, and the particles were redispersed by sonication. The
particle size is in the range of 8 to 10nm as characterised by transmission electron
microscopy (TEM). Water was used to adjust the final solid concentration to ca.
6% (wt) and carbowax 20M (Merck) was added to the colloid in a proportion of 1 -
50% of the TiO2 weight. The colloid thus obtained was used for the preparation
T/O2 nanoporous film electrode.
A2) A 16.0 cm3 of isopropanol (Aldrich, AR grade) and 50.0 cm3 of titanium
butoxide (Aldrich, AR grade) were accurately measured into a 150 cm3 dropping
funnel. The resulting solution was added over 15 minutes with vigorous stirring to
600 cm3 of ultrapure deionized water (18.2 M? cm) in a conical flask. On the
completion of the addition, 4.0 cm3 of 70% nitric acid (AR grade) was added into
the solution as a peptizing agent. The solution was immersed in a hot water
bath, heated to 80°C and stirred continuously for 10 hours. Approximately 400
cm3 of a white colloidal solution remained and was stored in a dark glass vessel for
use. The sizes of the TiO2 synthesised according to this procedure were in a
range of 8 to 10 nm.
The colloidal TiO2 prepared above was placed in an autoclave reactor (Parr
bomb) and autociaved for 12 hours at 200°C before concentrating on a rotary
evaporator to 8% (w/w), resulting in a white semi-viscous colloidal solution. 40%
TiC>2 weight equivalent (e.g. 1.6g in 50.0 cm3 of 8% colloidal solution) of Carbowax
6,000 was added to the solution and stirred for approximately 20 minutes.
B) Immobilisation of TiO2 film on ITO glass
B1) ITO (tin doped iridium oxide) conducting glass slides were used as the
substrate for immobilisation of TiO2 particles. To get a clean surface the ITO glass
slide was pretreated by washing in turn with detergent, water, and ethanol. After
the pretreatment the ITO slide was dip-coated in the above colloidal solution from
Al, above. The coated slides were then calcined in a muffle furnace in air at 500 to
800°C for 0.5h to 30h. The-particle size of TiO2 on the films characterised by x-ray
diffraction and scanning electron microscopy (SEA/I) is in the range of 10nm to
100nm and the rutile/anatase phase ratio is in the range of 0.1 to 50%.
B2) TiOz films were prepared in a clean room environment to minimize
contamination from dust particles. TiO2 colloidal coating solution, from A2 above,
was stirred vigorously and subjected to the ultrasonic treatment for 20 minutes
prior to a dtp coating process to achieve a consistent, reproducible homogeneous
mixture. The ITO slide (conducting glass) was used as the electrode substrate
and was pre-treated by washing in turn with detergent, water, acetone and water,
and finally dried by pure nitrogen. After pre-treatment, the VTO slide was dip-
coated with the TiO2 colloidal coating solution using a dip coating equipment with
withdrawing speeds of 0.5 - 1.5 cm/min. The coated electrodes were then
calcined in a muffle furnace at 450°C for 30 minutes in air. The nanoporous TiO2
films with I^rn thickness and anatase crystalline structure were obtained. The
films with different thicknesses can be prepared by controlling the withdrawing
speed during the dip coating.
C) General Setup of the Photoelectrochemical System
Fig. 2 shows the schematic of the instrumental set up of the
photoelectrochemical detection system. Illumination was carried out using a 150W
xenon arc /amp fight source with focusing lenses (HF-200W-95, Beijing Optical
Instruments). To avoid the sample solution being heated-up by infrared light, the
light beam was passed an UV-band pass filter (UG 5, Avotronics Pty, Limited) prior
to illumination of the electrode surface. A light shutter was used to control the ON
and OFF of the illumination.
Generally, photoelectrochemical experiments were performed in a three-
electrode electrochemical cell with a quartz window for illumination. The TiO2 film
electrode was installed in an electrode holder with ca. 0.65cm2 left unsealed to be
exposed to the solution for illumination and photoelectrochemical reaction. A
saturated Ag/AgCI electrode and a platinum mesh were used as the reference and
counter electrodes respectively. A voltammograph (CV-27, BAS) was used for
application of potential bias in steady state photocurrent measurements. Potential
and current signals were recorded using a Macintosh computer (7220/200)
coupled to a Maclab 400 interface (AD Instruments),
D) Th© measurement procedures:
D1) Exhaustive degradation conditions
Step 1: Once the system is set up (see Fig 2 - 6), the supporting electrolyte
solution was pumped through the photoelectrochemical cell (a thin layer cell). A
bias potential of +0.20 Vvs a Ag/AgCI reference electrode was applied. Once the
stable baseline was obtained, the pump was stopped. The photo shutter was then
switched on to allow UV radiation to reach the electrode. The photocatalytic
reaction occurred and the background current-time profile can be measured (see
Fig 10 curve (a)). The background current (iBackground) was resulted from the
photocatalytic oxidation of water. Integrating (Background with time we can obtain the
background charge, QBachkround.
Step 2: After the measurement of QBachkround. the photo shutter was switched
off and a 5µl to 200µI of sample solution with appropriate concentration was
injected into the photoeiectrochemical cell (It is to note that the sample volume
injected is depending the volume of the cell and if the concentration of organics in
the sample was too high then an appropriate dilution may be required prior the
injection). Once the sample injection was completed, the pump was stopped and
the photo shutter was switched on. The Current-time profile was measured (see
Fig 10 curve (b)). The current obtained here is the total photocurrent (iTotal) that
resulted from the oxidation of water and organics. Same to the above, by
integrating itotal with time we have the total charge, Qrota.-
Step 3: Since the background charge, QBackgvund, is a constant for the given
experimental conditions and the total charge, QTotal, varied with the concentration of
the sample, therefore, the net charge, Qnet (the shaded area shown in Fig 10) that
resulted from the oxidation of organics can be obtained by subtracting the
background charge from the total charge, that is:
Qnet= QTotal- QBackground
The COD value of the sample can then be calculated according to the equation (7)
since in the equation,

F is a constant and V is known sample volume.
Repeating steps 2 and 3 to analysis next sampie.
D2) Non-exhaustive degradation conditions
The system set up was same as described above (see Fig 2-6) except that a
normal flow-through cell was employed to replace the thin layer cell. The
measurement can be done by using a standard addition method or by other
calibration means.
Step 1: The supporting electrolyte solution was pumped through the
photoelectrochemica! cell (a normal flow-through cell with cell volume of 0.5 to 2.5
ml). A bias potential of +0.20 V vs a Ag/AgCl reference electrode was applied.
Once the stable baseline was obtained, the pump was stoped. The photo shutter
was then switched on to allow UV radiation to reach the electrode. The
photocatalytic reaction occurred and the background current-time profile can be
recorded (see Fig 7 curve (a)). The background current (ibackground) was resulted
from the photocatalytic oxidation of water.
Step 2; After the measurement of ibackground, the photo shutter was switched
off and a 0.5ml to 2.5ml of sample solution with appropriate concentration was
injected into the photoelectrochemica! cell (It is to note that the sample volume
injected is depending the volume of the cell and if the concentration of organics in
the sample was too high then an appropriate dilution may be required prior the
injection). Once the sample injection was completed, the pump was stopped and
the photo shutter was switched on. The Current-time profile was recorded (see Fig
7 curve (b)). The current obtained here is the total photocurrent (iTotal) that resulted
from the oxidation of water and organics.
Step 3: Once the measurement of iTotal was completed, the photo shutter was
switched off and a 0.5ml to 2.5ml of sample solution containing an appropriate
concentration of standard was injected into the photoelectrochemical cell. Once
the sample injection was completed, the pump was stopped and the photo shutter
was switched on. The current-time profile was recorded (see Fig 7 curve (c)). The
current obtained here is the photocurrent (istandard) that resulted from the oxidation
of water and the organics in both original sample and the added standard.
Step 4: After the above measurements, the COD of the sample can be
calculated according to the equation (10).
The limiting current for each case (/l background, il(1) and il(2)), can be obtained
by measuring the steady current value from each curve, for example, at 45s (see
Figure 5). The net limiting photocurrents for the sample solution (/;(1)) and for the
sample with added standard (/,(2)) can then be calculated.
Net limiting photocurrents (or current) for the sample solution:

Net limiting photocurrents (or current) for the sample with standard:

According to equation (10), above, we have:
Repeat the steps 2 to 4 for the next sample.
It will be appreciated by the person skilled in the art that the necessary
computations set out above may be automated with the appropriate programming
of a personal computer.
There are a few operational modes with different photoelectrochemica)
reactor designs (i.e. online thin-layer flow cell, and batch cell) that utilise the assay
methodology and are demonstrated by following examples.
Example 1: Quantification of COD using photocurreni
The photoelectrochemical experiment was performed in a three-electrode
electrochemical batch cell with a quartz window for illumination as shown in Fig. 3
The TiO2 film electrode was placed in an electrode holder with ca. 0.65cm2 left
unsealed to be exposed to the solution for illumination and photoelectrochemical
reaction. 0.1M NaNO3 solution was used as the supporting electrolyte. A potential
bias of +0.2V was applied at the electrode and limiting photocurrents were
obtained for different organic compound concentrations when the current reached
steady state. The limiting photocurrent differences between samples and the
blank 0.1M NaNO3 solution were taken as analytical signals, which are directly
linear to organic compound concentrations within diffusion control, A linear
relationship between the analytical signal and COD value was then acquired after
the concentration was converted into COD value.
Example 2. Quantification of COD using charges
In this case the experiment was carried out in a thin-layer
photoelectrochemical cell as shown in Figures 4 and 5. A potential bias of
+0.20V was applied and 2M NaNO3 was used as supporting electrolyte. Firstly, a
2M NaNO3 electrolyte solution was injected into the thin-layer
photoelectrochemica! cell with a syringe and a blank transient photoelectrolysis
was run as a blank sample. The photocurrent-time profile was recorded until the
photocurrent reached steady state. Then samples containing organic compounds
and 2M NaNO3 were injected into the thin-layer cell and the sample transient
photoelectrolysis was run. The photocurrent-time profile was recorded until the
photocurrent attained steady state, indicating the organic compounds have been
exhaustively photoelectrolysed. The cell was washed with supporting electrolyte
solution between each sample injection. Integrating the photocurrent-time profile
gives the photocatalytic oxidation charge. The charge difference between sample
and blank transient photoelectrolysis was taken as the analytical signal, which is
directly proportional to the COD value. COD value was then determined.
Example 3. Quantification of COD using charges and FIA
Besides the use of the thin-layer photoelectrochemicai cell, a flow injection
analysis (FIA) system was incorporated into the COD determination. With the
combination of FIA, automatic COD determination was realised. In this case, the
injection of samples and cell cleaning was controlled by a FIA controlling system
as shown in Fig. 6 (a). Pump 1 achieves the blank sample (R1) injection and cell
cleaning while Pump 2 does the sample injection (R2). A potential bias of +0.20V
was applied and 2M NaNO3 was used as supporting electrolyte (blank sample). .
Firstly, a 2M NaNO3 electrolyte solution was pumped into the thin-layer
photoelectrochemical cell by Pump 1 and a blank transient photoelectrolysis was
run as a blank sample. The photocurrent-time profile was recorded until the
photocurrent reached steady state. Then samples containing organic compounds
and 2M NaNO3 were pumped into the thin-layer cell by Pump 2 and the sample
transient photoelectrolysis was run. The photocurrent-time profile was recorded
until the photocurrent attained steady state, indicating the organic compounds
have been exhaustively photoelectrolysed. The cell was washed with supporting
electrolyte solution by Pump 1 between each sample. Integrating the
photocurrent-time profile gives the photocatalytic oxidation charge. The charge.
difference between sample and blank transient photoelectrolysis was taken as the
analytical signal, which is directly proportional to the COD value. COD
value was then determined.
Example 4. Quantification of COD using continuous flow mode
Besides the use of the thin-layer photoelectrochemicai cell, a flow injection
analysis (FIA) system was incorporated into the COD determination with a
continuous flow operational mode. With the combination of FIA, automatic COD
determination was realised. In this case, the injection of samples and cell cleaning
was controlled by a FIA controlling system as shown in Fig. 6(b). The blank
sample is continuously pumping through the cell and an injector is employed for
the sample injection. A potential bias of +0.20V was applied and 2M NaNO3 was
used as supporting electrolyte (blank sample). Recording the photocurrent from
the photocatalysis of the blank sample gives a steady baseline. An injection of
sample containing organic compounds through the injector to allow the
photocatalysis of sample takes place. The peak shaped photocurrent-time profile
can be recorded until the photocurrent attained baseline, indicating the organic
compounds have been photoefecfrolysed. The next samp/e can then be injected
for analysis. GOD value of the sample can be determined by measuring either
peak height or peak area (by integrating the peak photocurrent) since both peak
height and peak area are directly proportional to the COD value.
Typical experimental results
Fig 8 shows a typical photocurrent-time curve obtained from a non-
exhaustive photoeiectrochemica! degradation process. Under a constant applied
potential, when the light was switched off, the residual current (dark current) was
approximately zero. Upon illumination, the current increased rapidly before
decaying to a steady state value. This steady state current (limiting photocurrent)
consists of two current components. One is due to mass transfer limited
photoelectrochemicai oxidation (degradation) of the target analyte, which is directly
proportional to the concentration of the analyte. The other is due to the
photoelectrochemicai oxidation (decomposition) of water, which is constant at a
given pH and supporting electrolyte concentration. The net limiting photocurrent,
il, (resulting from analyte oxidation) can be readily obtained by subtraction of the
photocurrent attributed to the oxidation of water from the total photocurrent.
can then be used for analytical purposes (Eqn 10).
Preliminary results obtained from a range of organic compounds indicate
that Equation 10 is applicable to all compounds investigated (see Fig. 9, in which 1
= p-chlorophenol; 2 = potassium hydrogen phthalate; 3 = methanol; 4 = d-giucose;
5 = malonic acid; 6 = succinic acid; 7 = glutaric acid; 8 = glycine). As predicted, //
was found to be directly proportional to the concentration of the organic compound.
The slopes of the il-C curves (sensitivity) are determined by the mass transfer
coefficient (k) and the number of electrons transferred (n) during the
photoelectrochemical degradation.
The photocurrent-time profile of an exhaustive photoelectrochemical
degradation process was found to be similar to that of the non-exhaustive
degradation process except that a steady state photocurrent can only be achieved
when all of the organic compounds were consumed. In this case, the steady state
photocurrent was purely due to the oxidation of water and can be easily subtracted
from the total current (Fig. 10). Figure 10 shows the typical photocurrent - time
profiles obtained from the electrodes during an exhaustive photoelectrochemical
degradation process in phthalic acid and in blank electrolyte solutions. It can be
seen that the photocurrent decayed with time and then reached a steady value,
which was due to the oxidation of water. It is noted that the blank photocurrent
obtained from the blank electrolyte solution was purely due to the oxidation of
water, while the photocurrent obtained from the electrode in phthalic acid solution
consists of two current components, one is due to photoelectrochemical oxidation
of phthalic acid, and the other is due to the oxidation of water, which is the same
as the blank photocurrent. Our experimental results showed that the blank
photocurrent was essentially constant for the given set of experimental conditions.
For a given time period, the charge passed for both cases can be obtained by
integration of the photocurrent and blank photocurrent. The charge difference
between the two cases is the net charge, Q, due to the photoelectrochemical
oxidation of phthalic acid, which is indicated as the shaded area in Fig 10. The net
charge, Q, was measured by integration of net photocurrent within the degradation
period as described in Fig 10. As predicted by Eqn. 3, Q is directly proportional to
concentration (see Fig. 11). In this case, the slopes of the Q-C curves
(sensitivity) were dependent only on the number of electrons transferred (n). The
results in Fig. 9 and 11 (having conformed to theory) demonstrated the possibility
of further developing the proposed system into a "universal" detection system for
individual analytes.
The possibility of applying the proposed method for determining aggregate
properties such as COD was also tested. We chose the APHA COD standard test
solution (potassium hydrogen phthalate (KHP)) and a synthetic mixture with known
COD values as our test solutions. Fig 12 shows the correlation between the
experimental COD values (according to equation 7) and standard COD values.
Excellent agreements between the two COD values were obtained in both cases.
Example 5
Real Waste water Samples.
Fourteen (14) different wastewater samples were collected from various
industries in Queensland, Australia. After appropriate dilution, all samples were
subject to the COD analysis using our method and the standard COD method.
The COD values obtained from the two methods for all samples were then
correlated and shown in Fig 13. A correlation coefficient of 0.973 and slope of
0.992 were obtained. This means our method predicts the same COD value as the
standard COD method. This demonstrates that our method is equivalent to the
standard method in predicting the COD values.
Example 6
Determination of COD in Synthetic samples
The use of flow injection (Figures 4 and 5) to determine COD in aqueous solution
was first tested with synthetic samples prepared with pure organic chemicals, i.e.
glucose, glutamic acid, GGA and sucrose. The time required for a single
measurement was 1-2 min. Figure 14 shows the calibration curve of the various
organic compounds in terms of µM and theoretical COD concentration
respectively. Figure 14a shows that the photoelectrochemical detector had
different sensitivities (slope of the calibration curve) to different organics in regards
of µM concentration. The sensitivity decreased in the order of sucrose, GGA,
glucose and glutamic acid. This is because the organic compounds contribute
different number of electrons [n=4y- 2j+m-3k-q) in the exhaustive oxidation
reactions. With the decrease of transferred electrons per mole, i.e. sucrose
(n=48), GGA (n=42), glucose (n=24) and glutamic acid (n=18), the organic
compounds give fewer electrons per mole and hence the sensitivity decreases.
This explained the sensitivity order in Figure 14a. With the transferred electron
number (n), the concentrations of the organic compounds were converted from µM
to theoretical COD value in ppm. The same sensitivities, evidenced with the same
slope, were obtained for the selected organics in Figure 14b. This implied that the
photoelectrochemical detector oxidised the above organics to the same extent, i.e.
the organic compounds have been oxidised indiscriminately and the mineralisation
was achieved. The detection principle was therefore validated. These is shown by
plotting the PECOD values against the theoretical COD values as shown in Figure
14c using glucose trendline as standard calibration curve. The line of best fit has a
slope of 1.0268 and R2 of 0.9984, which directly demonstrated that suitability to
use glucose as a calibration standard to determine COD value for the unknown
sample.
It was found that the detection limit of 0.5 ppm COD with a linear range up to 60
ppm COD can be achieved under the experimental conditions employed using
glucose as testing anaiyte. The detection limit can be extended further by
increasing the sample injection volume while the linear range can be increased by
a further smaller injection volume.
Reproducibility and stability are important parameters for the usefulness of the
detector. The response reproducibility of the sensor to 100 µM glucose was
studied using repeated determinations (n=12) and RSD% was found to be 0.8%.
The detector is relatively stable. Significant baseline shift was observed for the
first two hours when the electrode was brand new due to some of the active TiO2
particles were not attached on the electrode surface enough firmly and was
removed by the carrier. The baseline became almost constant after these non-
stable active sites were removed. In fact all the data reported in this paper was
obtained from the same TiOz electrode. The electrode had experience the change
of pH (from 2 to10), the change of potential (-0.4 to +0.8V), the change of flow
rate, the change of injection volume and analysis of real samples and finished
nearly thousand of the measurements it is stiii relatively sensitive and stable.
When the electrode was not being used, it is filled with Milli-Q water and store in
the light. It is well known that TiO2 surface has merits of self-cleaning and super
hydrophilicity. The fouling of electrode, which is commonly caused by adsorption
of organic compounds and growth of bacteria, was not observed after storage.
Because of this, even after a few days, it needed only about 5 minutes to
regenerate the used electrode to acquire a stable baseline to start the detection of
COD again.
Example 7
PECOD vs COD
The pH of the real samples tested in this example was in the range of 5-9, which is
the pH independent region of the photoelectrochemical detector. Standard
calibration curve method was used to determine the COD value in real sample.
Figure 15 shows the typical response of the flow injection response using glucose
as the standard substance. As shown in the figure, both the charge (peak area,
Fig 15 (a)) and the peak current (peak height, Fig 15 (b)) increased proportionally
with the increase of glucose concentration. The calibration curve (the inserts of
Figure 15) was therefore constructed using the data from the above detection.
At the same time, the standard COD value was determined with conventional COD
method (dichromate method). Figure 16 shows the correlation between the
experimental COD values and standard COD values. Where valid, the Pearson
Correlation coefficient was used as a measure of the intensity of association
between the values obtained from the flow injection photoelectrochemical COD
method and the conventional COD method. This was employed for the data in
Figure 16. A highly significant correlation (r=0.991, P=0.000, n=13) between the
two methods was obtained indicating the two methods agreed very well. More
importantly, the slope of the principle axis of the correlation ellipse of 1.0819 was
obtained. This almost unity slope value suggests both methods were accurately
measuring the same COD value. Given a 95% confidence interval, this slope was
between 1.016 and 1.174. This implies that we can be 95% confident that the
true slope lies between these two values. Consider that there are analytical errors
associated with both the flow injection photoelectrochemical COD and the
standard method measurements and that these errors contribute to scatter on both
axes, the strong correlation and slope obtained provides compelling support for the
suitability of the flow injection photoelectrochemical COD method for measuring
Chemical oxygen demand. The excellent agreements between the two COD
values demonstrates the suitability for the proposed method to measure COD.
Conclusion
The present invention provides a COD analysis method, which is accurate,
sensitive, environmentally friendly, robust, rapid and easy to be automated. This
method in principle measures the theoretical COD value due to the extraordinary
high oxidation power of photogenerated holes. The method described here is a
direct method and independent of sample matrix. Under exhaustive degradation
conditions, the method is an absolute method requires no calibration.
Experimentally, it correlates well with the conventional dichromate method
(Standard method). The electrode has a very good long time-stability, without
showing any decline of photocatalytic activity. The nature of the analytical principle
employed makes the method insensitive to the change of temperature. The
method shows a good tolerance to temperature change in contrast to Karube's
method. During the experiment the temperature did not controlled with the sample
temperature ranging from 10 to 40°C, no photocurrent and charge change was
observed.
Whilst the description of the invention utilises a three electrode
photochemical cell it will be appreciated that the photoetectrochemica! cell may
comprise a working electrode (TiO2) and a counter electrode, wherein the counter
electrode may suitable act as both a counter electrode and a reference electrode.
Throughout the specification the aim has been to describe the preferred
embodiments of the invention without limiting the invention to any one embodiment
or specific collection of features.
Throughout this specification, unless the context requires otherwise, the
word "comprises", and variations such as "comprise" or "comprising", will be
understood to imply the inclusion of a stated integer or group of integers or
steps but not to the exclusion of any other integer or group of integers.
CLAIMS
1. A method of determining chemical oxygen demand of a water sample,
comprising the steps of
a) applying a constant potential bias to a photoelectrochemical cell, having
a photoactive working electrode and a counter electrode, and containing
a supporting electrolyte solution;
b) illuminating the working electrode with a light source and recording the
background photocurrent produced at the working eiectrode from the
supporting electrolyte solution;
c) adding a water sample, to be analysed, to the photoelectrochemical cell;
d) illuminating the working electrode with a light source and recording the
total photocurrent produced with the sample;
e) determining the chemical oxygen demand of the water sample
according to the type of degradation conditions employed.
2. A method as claimed in claim 1 wherein the photoactive working
electrode is a nanoparticulate serniconductive electrode.
3. A method as claimed in claim 2 in which the working eiectrode is a iayer
of titanium dioxide nanoparticles coated on an inert conductive
substrate.
4. A method as claimed in claim 1 in which a reference electrode is also
used in addition to the working and counter electrodes.
5. A method as claimed in claim 1 in which the chemical oxygen demand is
determined under exhaustive degradation conditions, in which a!i
organics present in the water sample are oxidised.
6. A method as claimed in claim 1 in which the chemica! oxygen demand is
determined under non-exhaustive degradation conditions, in which the
organics present in the water sample are partially oxidised.
7. A method as claimed in any preceding claim in which the background
photocurrent is deducted from the total photocurrent produced with the
sample to obtain the photocurrent due to the oxidation of organic
material in the sample.
8. A method as claimed in claim in any preceding ciaim in which the
sample is diluted with the supporting electrode.
9. A method as claimed in claim 1 in which the chemical oxygen demand is
determined by measuring charge or current under exhaustive
degradation conditions with a stationary or flow cell using different
operational modes including batch mode, flow-stopped mode and
continuous flow mode.
10. A method as claimed in claim 1 in which the chemical oxygen demand is
determined by measuring charge or current under non-exhaustive
degradation conditions with a stationary or flow cell using different
operational modes including batch mode, flow-stopped mode and
continuous flow mode.
11. A photoelectrochemical assay apparatus for determining oxygen
demand of a water sample which consists of
a) a measuring cell for holding a sample to be analysed
b) a photoactive working electrode and a counter electrode disposed in
said cell,
c) a light source adapted to illuminate the photoactive working electrode
d) control means to control the illumination of the working electrode, the
applied potential bias, and photocurrent recording
c) photocurrent/charge measuring means to measure the
photocurrent/charge at the working electrode
f) analysis means to derive a measure of oxygen demand from the
measurements made by the photocurrent/charge measuring means.
12. Apparatus as claimed in claim 11 in which the measuring cell is a flow
through cell.
13. Apparatus as claimed in claim 11 or 12 in which a reference electrode is
included in the measuring cell.
14. Apparatus as claimed in any one of claims 11 to 13 wherein the
photoactive working electrode is a nanoparticulate semiconductive
electrode.
15. Apparatus as claimed in any one of claims 11 to 14 in which the
working electrode is a layer of titanium dioxide nanoparticles on an inert
substrate.
16. Apparatus as claimed in any one of claims 11 to 15 which also includes
a reservoir for a supporting electrolyte which is used to measure the
background photocurrent and to dilute the sample.
17. Apparatus as claimed in claim 16 which also includes a sample
supply/injection system and a supporting electrolyte supply/injection
system
A method for
determining chemical oxygen demand
of a water sample comprises the steps
of (a) applying a constant potential
bias to a photoelectrochemical
cell, having a photoactive working
electrode (e.g. a layer of titanium
dioxide nanoparticles coated on an
inert conductive substrate) and a
counter electrode, and containing a
supporting electrolyte solution; (b)
illuminating the working electrode
with a light source and recording the
background photocurrent produced
at the working electrode from the
supporting electrolyte solution; (c)
adding a water sample, to be analysed,
to the photoelectrochemical cell; (d)
illuminating the working electrode
with a light source and recording the
total photoelectrocurrent produced
with the sample; (e) determining
the chemical oxygen demand
according to the type (exhaustive
or non-exhaustive) of degradation
conditions employed. An apparatus
for carrying out the method is also
claimed.

Documents:


Patent Number 223032
Indian Patent Application Number 01849/KOLNP/2005
PG Journal Number 36/2008
Publication Date 05-Sep-2008
Grant Date 03-Sep-2008
Date of Filing 16-Sep-2005
Name of Patentee AQUA DIAGNOSTIC PTY LTD
Applicant Address LEVEL 1, 159 DORCAS STREET, SOUTH MELBORNE, VIC
Inventors:
# Inventor's Name Inventor's Address
1 ZHAO, HUIJUN 34 KINCAID DRIVE, HIGHLAND PARK, QLD 4211
PCT International Classification Number G01N 27/30; G01N 33/18
PCT International Application Number PCT/AU2004/000438
PCT International Filing date 2004-04-05
PCT Conventions:
# PCT Application Number Date of Convention Priority Country
1 2003901589 2003-04-04 Australia