Title of Invention

A HEMODIALYSIS CARTRIDGE

Abstract The scope of the invention is to apply an electric current or to use u kind of iontophoresis system with the hemodialysis cartridge and system (the proposed method is also applicable to peritoneal or other similar methods) to remove unwanted molecules from blood, plasma or serum or other body fluids and to increase the effectiveness of the process. This cartridge can be used for patients with uremia and cartridge fixed to the conventional hemodialysis machine and additionally the electric current applied to the electrodes placed in to the cartridge or electrode connectors placed to the conventional cartridge. When the system activated, the molecules in the blood or other body fluid migrates to the hemodialysis solution. Charged ions or uncharged molecules move together with electrosmotic flow. The sterilized electrodes preferably made by Ag/AgCl to prevent pH changing effect. Other apparatus can also be used for providing an electropotential gradient.
Full Text A HEMODIALYSIS CARTRIDGE
This invention is related to a cartridge and/or a method that provides an additional step of
iontophoresis to conventional/classic hemodialysis procedure for those patients who have
insufficient kidney functions. It enhances the hemodialysis performance in removing of urea from
the blood using electrical potentials. Similarly, it is also related with the procedures such as
peritoneal dialysis or related with the procedure when some compounds or molecules (charged or
uncharged atoms or molecules, elements or ions) need to be removed from the blood (blood,
plasma or serum) to dialysis solution in acute or chronic poisoning cases. It is possible to increase
the efficiency of the hemodialysis method and to reduce the total time period of the procedure
using this proposed cartridge and/or the method. There are no similar procedures and/or
cartridges in use so far.
Basically, hemodialysis is a process to remove urea and some other toxic compounds
from blood into the hemodialysis solution by passive diffusion. In this procedure, a semi-
permeable membrane is used as a dialysis membrane. While the blood is circulating
continuously at the one side of the hemodialysis membrane, the hemodialysis solution at the
other side, continuously circulates as well. During the process, urea present in the blood at high
concentration, depending on the concentration gradient, it passes through the membrane from
blood to the hemodialysis solution. Thus, the urea concentration in blood decreases by the time.
In conventional hemodialysis procedures, the patient is connected to the hemodialysis machine
for about 4 hours, and urea concentration generally decreases to 50% of the beginning level even
at the best circumstances.
In iontophoresis procedure, by using an electrical current (creating an electrical potential
difference), the ions (molecules or atoms that having a net charge or partially charged) can be
carried to the other side of the membrane according to applied current and electrical charge and it
is possible to control it. Electrodes or similar tools are provided to the both side of the membrane
and the applied electrical current or potential can vary as needed. The ions in the solution/blood
migrate according to their charges and their movement is in proportional to the current. For
instance positively charged ions migrate to the negative electrode side and vice versa. While the

charged ions are migrating according to the electrical current, they also drag the uncharged
molecules along with moving molecules. At this instance, traveling from one side to the other
side of the membrane creates a flow, a turbulence occurs (this is called an electro-osmotic flow).1
Therefore the unchanged particles (atoms/molecules) can also be able to pass through the
membrane by being pulled into this vortex or into the motion and, this passage occurs at a much
faster rate than that of passive diffusion.
When the molecular structure of urea investigated, it is seen that some local charges are
present on the molecule. According to the experiments we've conducted, the higher urea
transportation was observed than passive diffusion and cathodal iontophoresis when urea was
present at the positive electrode side because of the positive local charges on the molecule. There
is also a possibility that the electroosmotic current was partially influential for this transfer.
However, during the transfer, if the other small but charged ions like potassium and sodium are
present, the transfer rate decreases; but still the transfer is much larger and faster than the
classical method. These experiments were repeated using human blood obtained from the
patients with uremia and similar results were achieved. With this invention the hemodialysis
procedure is shortened in time and, simultaneously, provided much better result (cleaner blood).
The previously conducted diffusion experiments were repeated with a peristaltic pump
using human blood, hemodialysis solution and the hemodialysis cartridge in new proposed
design. In the analysis of the samples taken from the blood that went through the cartridge, it is
found that the urea level in blood when the iontophoresis procedure was used prompted 3 to 5
times faster rate of decrease than the classical hemodialysis results. In other words, while the
classical hemodialysis procedure takes 4 hours, the iontophoresis procedure of ours lowers the
process time to about 30 minutes, and with much better results (Figure, 1).
If the classical hemodialysis cartridge and proposed iontophoresis procedure is going to
be used together; system is depicted in Figure 2. The cartridge used as a hemodialysis cartridge
for removing urea from the blood simply comprises a holding compartment (Figure-2-1- and
Figure-3-1); and because of the electrodes placed in the cartridge, (Figure-2-A and B; Figure-3-A
and B) electric current can be applied and that, it is also designed for the use of classical
hemodialysis cartridge.

Accordingly, in this figure, the point labeled as 1 indicates the entry of blood and the label
2 shows the exit of the blood. The labels 3 and 4 show the entry and exit points of the
hemodialysis solution. The label 5 shows the hemodialysis cartridge's dialysis membrane. The
label B shows positive electrode, and label A represents the negative electrode. Thus, when the
system is activated, the blood and hemodialysis solution is circulating continuously and electric
current is applied and at the end urea can be transferred to the hemodialysis solution with much
faster rale. The labels A and B are the electrodes made by Ag and AgCl or they can be designed
for same purpose in different shape or compositions. If needed, it is possible to use Ag for B, and
AgCl for A.UV or ethylene oxide sterilization can be used for the sterilization of the electrodes.
On the other hand, although the composition of the sterilized electrodes (Figure-2-A and
B; Figure-3-A and B) is preferred to be Ag/AgCl for preventing the pH effect of the electrodes,
they can be also made out of platinum, copper, gold, steel, graphite, vanadium, tungsten, etc. The
composition, shape, design, connection point and their place in and outside of the cartridge, are
not deterministic and specific properties for the electrodes. Some apparatuses which can be used
in formation of electrical gradient may be utilized.
In the experiments, when the clectrohemodialysis procedure was used, the sodium and
potassium levels of the blood and the hemodialysis solutions have been analyzed. When this
procedure used, the level of sodium and potassium is lowered in blood as well. This outcome is
possibly same for some other ions and unchanged molecules. For preventing some possible
complications, the ions and other material in the hemodialysis solution need to be adjusted.
Hemodialysis solutions must be prepared according to the patients' needs. Or, the blood that
exited from the hemodialysis cartridge can be connected to another cartridge and similarly ions
can be replaced using reverse current and the problem can be solved. The hemodialysis
solution(s) (at the beginning and at the supplementary durations) can be prepared according to the
needed requirements of the patient to avoid any possible complications.
Additionally, in these complications (the problems like imbalance of electrolytes and/or
osmotic pressure or similar unwanted outcomes due to iontophoresis) problems can be overcame
by producing a membrane that would have smaller pores. With two membranes present in one,
first one allows the unwanted urea and the other molecules does not allowed by second material
to go through. The application of electrical current to both membrane or by applying the classical

methods to the second one (utilizing only the concentration gradient) the possible problems can
be avoided.
There is no research and experiments that have been published so far, the application of
electrical current or potential and the gradual electrical effect have not been tried for the
hemodialysis or peritoneal dialysis or removing unwanted molecules, ions etc. from the blood or
other body fluid using a system like our proposed system here. In this study, the addition of
iontophoresis procedure, in terms of shortening the time duration and bettering the quality of
outcome; is a revolutionary format that is at the cutting edge of the known medical technology.
In the literature there is no study or research have been conducted to this end.
This invention, in addition to its speed and quality in the treatment of uremia, it has a high
potential of removing unwanted/unneeded non-polar and especially polar substances during the
acute and chronic poisoning; similarly with electrical current and the cartridge usage the
unwanted items/substances in the blood can be pulled out into the hemodialysis solution. In the
literature we came across that the electrical current lesser than 0.5 mA/cm2 does not cause
damage to the blood or body cells1. For this reason the current that is low than 0.5 mA/cm2 would
be an acceptable for a positive out come. The magnitude of the current can be chosen in required
levels. For the system, direct or alternative current, square, sinus or triangular or even different
frequencies and/or currencies can be applied. In this system, the electrical flow/current does not
make any direct contact with any of the body cells; and therefore, it would be possible to exceed
the electrical level of the aforementioned literature. However, because there is a likelihood of
damaging the blood cells during a high currency flow this would not be recommended.
Additionally, the pH levels was not affected by the Ag/AgCl electrodes, therefore we
preferred. Electrodes can be made with different compounds or also some similar apparatuses
with similar functions can be used. The electrodes may even be simply attached to the classical
hemodialysis cartridges' blood and hem-dialysis solution's entry ports (Figure 3). If the available
classical hemodialysis cartridge will be used, an alternative placement of the electrodes are
shown in Figure 3. In this model, the electrodes can be attached to blood entry port and
hemodialysis solution part as in the previous cases, and the classical hemodialysis cartridge can
be used with minimal modifications. (In this model, electrodes in different composition or some

apparatus with similar functions can be used). The electrodes themselves need to be sterilized. In
accompanying figure 3. the labels 1,2,3,4 and 5 are the same labels 1,2,3,4 and 5 are the
same labels as shown in Figure 2.
As a result, this invention will prevent those patients who have insufficient kidney
functions being hooked up to a hemodialysis machine for a long time. It will provide for the
urea, creatine, and some toxic compounds to exit the blood in a better way. Additionally, the
procedure's potential of removing the unwanted elements from the patient's blood stream in
acute and chronic poisonings will provides a very valuable device in the field of medicine.
With the exception of electrodes and similar apparatuses, the cartridge being proposed, in
terms of shape and dimensions is very similar to chose cartridges used in classical hemodialysis.
The difference of the cartridge, such as surface area or its membrane with different pore sizes,
alone is not a distinctive quantification of the proposed device for patent purposes. The sizes,
compositions, or locations (such as one being by the blood inflow side and the other being by the
dialysis side or the solution) of the electrodes and' or some apparatuses with similar functions do
not restrict its applicability for patent rights.
Additionally, the composition of the hemodialysis solution, the properties of the dialysis
membrane (color, texture, latex or biological tissue, pore size, selectivity, etc.), the way, the flow
intensity and the direction of the blood, hemodialysis solution and the number of cartridge used
connected before or after each other are not also deterministic and specific properties of the
invention.
Reference
1-M. J. Pikal, The role of electro-osmotic flow in transdermal iontophoresis," Advanced Drug
Delivery Reviews, 46, 281-305,2001.

WE CLAIM:
1 A hemodialysis cartridge comprising a blood, plasma circulation chamber and a
hemodialysis solution circulation chamber, a semi-permeable membrane located between said
blood/plasma circulation chamber and said hemodialysis solution circulation chamber, a first
electrode located in the said blood plasma chamber or in the cartridge inlet and an oppositely
charged second electrode located in the said hemodialysis solution chamber or in the cartridge
inlet, the said cartridge being adapted to cause continuous circulation of the blood plasma and
the hemodialysis solution on activation of the electrodes for How of molecule or substances
from the said blood plasma chamber to the said hemodialysis chamber.
2. A hemodialysis cartridge as claimed in claim 1. wherein said first and second
electrodes are made of the material selected from the group: silver, silver chloride, platinum,
copper, gold, steel, graphite and wolfram.
3. A hemodialsis cartridge as claimed in claim 2. wherein said first electrode is a silver
electrode and the said second electrode is a silver chloride electrode.
4. A hemodialysis cartridge as claimed in claim 1 . wherein an additional membrane with
smaller pore size is used for preventing entry of smaller ions and uncharged molecule present
with urea into the said hemodialysis chamber.

5. A hemodialysis cartridge as claimed in any preceding claim wherein said first electrode
is positively charged while the said second electrode is negatively charged.
6. A hemodialysis cartridge as claimed in claim 1. wherein the current density used in the
cartridge is less than or equal to 0.5 A cm2.
7. A hemodialysis apparatus incorporating the hemodialysis cartridge as claimed in any of
the preceding claims.

The scope of the invention is to apply
an electric current or to use u kind of iontophoresis
system with the hemodialysis cartridge and system
(the proposed method is also applicable to peritoneal
or other similar methods) to remove unwanted
molecules from blood, plasma or serum or other body
fluids and to increase the effectiveness of the process.
This cartridge can be used for patients with uremia
and cartridge fixed to the conventional hemodialysis
machine and additionally the electric current applied
to the electrodes placed in to the cartridge or electrode
connectors placed to the conventional cartridge. When
the system activated, the molecules in the blood or
other body fluid migrates to the hemodialysis solution.
Charged ions or uncharged molecules move together
with electrosmotic flow. The sterilized electrodes
preferably made by Ag/AgCl to prevent pH changing
effect. Other apparatus can also be used for providing
an electropotential gradient.

Documents:

788-KOLNP-2004-(03-01-2012)-FORM-27.pdf

788-KOLNP-2004-CORRESPONDENCE.pdf

788-KOLNP-2004-FORM 27.pdf

788-kolnp-2004-granted-abstract.pdf

788-kolnp-2004-granted-claims.pdf

788-kolnp-2004-granted-correspondence.pdf

788-kolnp-2004-granted-description (complete).pdf

788-kolnp-2004-granted-drawings.pdf

788-kolnp-2004-granted-examination report.pdf

788-kolnp-2004-granted-form 1.pdf

788-kolnp-2004-granted-form 18.pdf

788-kolnp-2004-granted-form 2.pdf

788-kolnp-2004-granted-form 3.pdf

788-kolnp-2004-granted-form 5.pdf

788-kolnp-2004-granted-reply to examination report.pdf

788-kolnp-2004-granted-specification.pdf


Patent Number 227306
Indian Patent Application Number 788/KOLNP/2004
PG Journal Number 02/2009
Publication Date 09-Jan-2009
Grant Date 06-Jan-2009
Date of Filing 09-Jun-2004
Name of Patentee DIZAYN TEKNIK PLASTIK BORU VE ELEMANLARI SANAYI VE TICARET A.S.
Applicant Address HADIMKOY YOLU SAN BIR 1. BOLGE 4. CAD. NO:23 34860 BUYUKCEKMECE ISTANBUL
Inventors:
# Inventor's Name Inventor's Address
1 DEGIM Y. TUNCER GAZI UNIVERSITESI ECZACILIK FAK. FARMASÖTIK TEKNOLOJI ANA BILIM DALI, ETILER 06330 ANKARA
2 DUNDAROZ RUSEN BAGKUR BLOKLARI, 4. BLOCK 69/14 ETLIK 06010 ANKARA
3 DENLI METIN GENEL KURMAY SAGLIK KOM. BASKAN YRD./KIZILAY ANKARA
4 ILBASMIS SIBEL GAZI UNIVERSITESI ECZACILIK FAK. FARMASÖTIK TEKNOLOJI ANA BILIM DALI ETILER 06330 ANKARA
5 OZCELIKAY A TANJU ANKARA UNIVERSITESI ECZACILIK FAK. FARMASÖTIK TEKNOLOJI ANA BILIM DALI TANDOGAN 06330 ANKARA
6 DEGIM ZELIHAGÜL GAZI UNIVERSITESI ECZACILIK FAK. FARMASÖTIK TEKNOLOJI ANA BILIM DALI ETILER 06330 ANKARA
PCT International Classification Number A61M 1/16
PCT International Application Number PCT/TR02/00073
PCT International Filing date 2002-11-19
PCT Conventions:
# PCT Application Number Date of Convention Priority Country
1 2001/03317 2001-11-19 Turkey